Test Equipment...

Halide Test Equipment

NCS Model LMP-2000 New Model!

The LMP-2000 is a state-of-the-art digital instrument specifically developed for carbon in-place testing per ANSI/ASME N510.

 

Like the earlier generation LMP-10, the LMP-2000 is housed in a field-rugged aluminum case, which contains an on-board carrier gas cylinder. Some of the many updated features include an external cylinder shut-off valve, floppy disk drive, LCD data display, numeric keypad, and a microprocessor controlling all internal functions. Computer data output includes a chromatogram, raw data, upstream and downstream tracer gas concentrations, instrument calibration data, system identification, and final results. The detailed report requires only an authorized signature for completion.

Using on-board instructions, only minimal training is required for plant personnel to perform an in-place leak test on a carbon adsorbed bank. A Quick Start Guide is included that will have you up and running in a matter of minutes.

  • SUPERIOR TRACER GAS SENSITIVITY
  • DISTINGUISHES TRACER GAS FROM BACKGROUND
  • USABLE WITH R-11, R-112, AND PERFLUOROCARBONS (PDCB)
  • SINGLE DETECTOR MEASUREMENT CONFIGURATION
  • BASED ON A STANDARD ANALYTICAL TECHNIQUE

ONBOARD FEATURES:

  • Microprocessor control of data analysis and output
  • Data output via RS-232 to PC for data storage and report generation
  • Floppy disk data storage
  • Direct data analysis and report generation capability
  • Chromatogram and report output via printer and/or PC

The LMP-2000 is an easy test instrument to master. The LMP-2000 automatically identifies tracer gas desorption resulting from wet or degraded carbon and calculates the system leak rate from a best fit, non-linear plot to time zero. This feature takes the guess work out of testing when working with difficult systems that exhibit rapid tracer gas desorption.

FLOWRATE ACCESSORY PACKAGE

The optional flowrate accessory package for the LMP-2000 allows the user to perform duct and stack flowrate measurements using the methodology specified in ASTM E2029 "Standard Method for Volumetric and Mass Flow Rate Measurement using Tracer Gas Dilution". In this standard test method, tracer gas is injected into a duct or stack at a known rate and the resulting concentration in the duct measured.

Using the flowrate accessory package, the LMP-2000 will monitor the tracer gas injection rate, automatically sample for tracer gas concentration, and calculate a duct flowrate. In many cases duct or stack flowrate data can be obtained more quickly and more accurately using the LMP-2000 in conjunction with the flowrate accessory than with conventional hot wire or Pitot tube traverse methods.

NCS Model LMP-6

  • Portable Self Contained
  • Minimum operator training
  • Stable refrigerant output
  • External compressed gas not required

SPECIFICATIONS:

Maximum metered R-11 output: 52 gm/min
Reservoir capacity: Nominal 1 liter
Delivery pressure: 40-50 PSIG
Power requirements: 120±10%, VAC 50/60 Hz, 3 AMP
Weight: 23lbs.

Aerosol Test Equipment

ATI manufactures the smallest, lightest photometers available in portable self-contained units. Our photometers have been designed with the user in mind for simplicity of operation and virtually maintenance-free use. Both our digital and analog units are available in version suitable for nuclear applications. Optional scanning probe and rugged carrying case are also available.

2h

  • size   10.1" x 14.3" x 5.8" (25.7 cm x 36.3 cm x 14.7 cm)
  • weight   15.5 lbs. plus 2.5 lbs. for scanning probe and access. = 18.0 lbs (7.0 kg plus 1.1 kg for scanning probe and access = 8.2kg)
  • enclosure   Instrument enclosures are rugged, die cast aluminum and have a swiveling handle that also acts as a support.
  • power   100 to 250 volts AC, 50/60 Hz fused power entry module. Automatically adjusts to all standard voltages used worldwide. Please specify power cable requirements when ordering unit.
  • auto-ranging   Provides 6 decodes of display from 0.000.1% to 100.0%, eliminating the manual range selection switch.
  • auto zero   Automatically establishes zero when sampling the airflow through the internal ULPA reference filter or when sampling the downstream airflow by pressing the 0% button.
  • alarm   larm feature sounds, if enabled, and causes main & probe displays to flash when user adjustable set point is exceeded.
  • bargraph   Solid-state bargraph with high visibility LED's on control panel and scanning probe.
  • amplifier   Solid-state linear amplifier.
  • photometer accuracy   1% of full scale for the decode in use (ratio metric operation).
  • digital display   Shows % Leakage from 0.000.1% to 100.0%
  • dynamic range   0.0001 to 125 micrograms per liter. Improved firmware prevents system errors by blanking the display when sampling above 125 micrograms per liter; display is restored when concentration decreases. "HIGH" mode enables sampling of concentrations exceeding 200 micrograms per liter.
  • light source   Stable, solid-state scattering chamber light source allows operation of lower power, generating less heat, and reduced stray light values. Life expectancy is 50,000 to 100,000 hours.
  • repeatability   0.5% full scale
  • scattering chamber   Field-tested design provides efficient and reliable performance. A three-stage "OPTICS" indicator ensures accurate internal reference response; Green = +/- 5% of original value, Orange = +/- 5-10% of original value, Red = > +/- 10% of original value.
  • vaccuum pump   1 cubic foot (28.3 liter) per minute sample rate, using a twin-head, carbon vane design, for quiet operation and dependability. Attached, external HEPA filter captures any expelled particulate. "FLOW" indicator ensures operation within required parameters.
  • output   a DB9 connector is provided standard on all units for RS232 output.

2ha

  • size   10.1" x 14.3" x 5.8" (25.7 cm x 36.3 cm x 14.7 cm)
  • weight   15.5 lbs. plus 2.5 lbs. for scanning probe and accessories = 18.0 lbs (7.0 kg plus 1.1kg for scanning probe and accessories = 8.1 kg)
  • enclosure   Instrument enclosures are rugged, die cast aluminum and have a swiveling handle that also acts as a support.
  • power   100 to 250 volts AC, 50/60 Hz fused power entry module with spare fuses. Automatically adjust to virtually all standard voltages used worldwide. Please specify power supply cable when ordering unit (UL/CS/VDE).
  • alarm   Audible Alarm in unit sounds when adjustable set point is exceeded. LED on probe lights up when set point is exceeded.
  • amplifier   Solid-state linear amplifier.
  • photometer accuracy   +/- 2% of full scale for the decade in use.
  • dynamic range   0.0001 to 125 micrograms per liter.
  • light source   Improved, longer lasting and more stable solid-state scattering chamber light source. Photometer operates at lower power generating less heat and lower stray light values. Life expectancy is 50,000 to 100,000 hours.
  • repeatability   1.0% full scale
  • scattering chamber   Redesigned smaller, lighter and more efficient for improved performance. Easier to clean and requires less maintenance.
  • vacuum pump   True 1 cubic foot per minute (cfm) sample rate (28.3 liters per minute) using a twin head, carbon vane pump for quiet operation and attached to the outlet of the vacuum pump to entrap any particles from the pump exhaust.
  • output   A DB9 connector is provided standard on all units for a 0-1 VDC output.

ATI manufactures portable Laskin nozzle aerosol generators which produce a submicron polydispersed aerosol in concentrations from 10 to 100 micrograms per liter at air flows from 50 to 8,100cfm @ 20 psig. All three Laskin nozzle generators feature rugged stainles steel construction, large fill ports, individual nozzle control and a 3" standard sanitary fitting outlet. An optional adapter is available for introducing the aerosol into positive pressure systems. Whatever your testing needs, we have a generator that will work for your application.


tda-4b lite

  • aerosol output range   50-4,000 cfm
  • aerosol concentration 100 ug/1 @*:   405 cfm
  • aerosol concentration 10 ug/1 @*:   4,050 cfm
  • generator type   1 to 3 laskin nozzles
  • compressed air   3 to 9 cfm @ 20 psi
  • aerosol type   poly-dispersed (Cold)
  • size   10" L x 8" W x 9" H (25 cm L x 20 cm W x 23 cm H)
  • weight   12 lbs. (5.5 kg)
  • electrical   (not required)

tda-4b

  • aerosol output range   50-8,100 cfm
  • aerosol concentration 100 ug/1 @*:   810 cfm
  • aerosol concentration 10 ug/1 @*:   8,100 cfm
  • generator type   1 to 6 laskin nozzles
  • compressed air   3 to 18 cfm @ 20 psi
  • aerosol type   poly-dispersed (Cold)
  • size   10" L x 11" W x 9" H (25 cm L x 28 cm W x 23 cm H)
  • weight   16 lbs. (7.3 kg)
  • electrical   (not required)

tda-6c

  • aerosol output range   50-2,000 cfm
  • aerosol concentration 100 ug/1 @*:   200 cfm
  • aerosol concentration 10 ug/1 @*:   2,000 cfm
  • generator type   2 to 6 laskin nozzle jets
  • compressed air   (not required, unit has built-in compressor)
  • aerosol type   poly-dispersed (Cold)
  • size   21" L x 11" W x 8" H (54 cm L x 28 cm W x 20 cm H)
  • weight   55 lbs. (25 kg)
  • electrical   120 VAC, 60HZ or 220V, 50HZ

tda-5b

TDA-5B produces a sub micron polydispersed aerosol in concentrations from 10 to 100 micrograms per liter at airflows from 1,500 to 65,000 cfm. The TDA-5B is recommended for certification applications where the required aerosol volume or concentration may vary widely from high to low (such as in large cleanroom plenums or individual filter units, where compressed air is not available, or portability necessary). An optional adapter is available for directing aerosol to an introduction point operating at less than ambient atmospheric pressure.

  • aerosol output range   1,500-65,000 cfm
  • aerosol concentration 100 ug/1 @*:   6,500 cfm
  • aerosol concentration 10 ug/1 @*:   65,000 cfm
  • generator type   heated chamber
  • compressed air   not required
  • compressed gas (inerf)*   20 cfh @ 50 psi
  • aerosol type   poly-dispersed
  • size   14" L x 10" W x 10" H (35 cm L x 25 cm W x 25 cm H)
  • weight   20 lbs. (25 kg)
  • electrical   110 VAC / 60HZ or 220V / 50HZ

*For high volume output, more uniform particle size, and safety, the liquid is aerosolized, superheated to vapor then re-condensed by ambient air into a poly-dispersed aerosol. The propellant must be an inert gas since the liquid vapor temperature is above the liquid flashpoint.

Test Reagents

  • PAO (Emery 3004/PAO-4)
    available in 1 and 5 gallon containers, MSDS available upon request
    CAS # 68649-12-7
     
  • DOP (DEHP)
    available in 1 and 5 gallon containers, MSDS available upon request
    CAS # 117-81-7
     
  • PSL (Poly Styrene Latex Spheres)
    available in 1 gallon containers, MSDS available upon request, containers of pre-mixed, near mono-dispersed solution in the following sizes:
    0.13 micron, 0.178 micron, 0.22 micron, 0.33 micron, 0.54 micron